Device not ready for indentation

Error Code 405 - DEVICE_NOT_READY_FOR_INDENTATION

Meaning:

Triggered if user tried to start a test or some other operation before all devices are ready. Different specific errors might be triggered depending on the device.

 

Troubleshooting Steps: 

Depending on the device(s) that have the issue.
To see the devices, go to 'View' > 'Show Connection Status'

  • Actuator motor is turned off:
    • Go to Tools -> Enable motor
  • Stage module is not parked - cannot start the indentation:
    • Park the profilometer - found in the 'Tools' tab
  • Actuator, DAQ or Stage module are not ready:
    • Close CORSICA and turn off the PLX-Benchtop
    • Unplug and Replug all the cables from the PLX-Benchtop
    • Check the emergency stop is not activated 
    • Restart your PC
    • Turn on the PLX-Benchtop and check that power is going to the machine  

    Contact support if the issue persists