Load cell reading too high

Error Code 110 - LOAD_CELL__READING_TOO_HIGH

Meaning: 

Load cell reading is higher than allowed
(the allowable value will depend on the operation taking place)
 
Troubleshooting Steps: 
  • If it happens when trying to start an indent or profilometry:
    'Balance' the load cell at rest and retry
  • If it happens when the indentation test is taking place, this means that the indent was trying to achieve a load > 9000 N:
    Review the indent set up (especially if using advanced parameters!)
  • If it happens at an intermediate indentation state (when the tip is not in contact with the sample):
    re-do the 'machine Set Up' in 'Tools' and check that you have the correct tip assigned in CORSICA.
  • If it happens when the profilometer is approaching the sample:
    'Balance' the profilometer at rest

Contact support if the issue persists