LVDT reading too high to start acquisition
Error Code 104 - LVDT__READING_TOO_HIGH_TO_START_EXPERIMENT
Meaning:
Triggered at the start of calibrations, indentation, and positioning for profilometry if the LVDT reading is not close to zero when you are about to start an operation.
Troubleshooting Steps:
- Check that the LVDT is not pressed when starting the test
(Move crosshead up if needed) - 'Balance' the LVDT while at its rest position so it is close to zero
- Check that, when gently pressed, the LVDT responds as expected
- Re-do the 'Machine Set Up' in the 'Tools' menu
Contact support if the issue persists.