Stage positioning error occurred. Please park profilometer and try again with the Take profile function under the Tools tab

Error Code 136 - STAGE_Y__POSITIONING_TIMEOUT

Meaning: 

One of the stages failed to execute the desired movement within the specified timeframe.

 

Troubleshooting Steps: 

  • Turn off CORSICA and the PLX-Benchtop
  • Hold the vertical arm holding the profilometer in place from behind and move the to and from the limit switches on the left and right side to check if there are no obstacles
    (Be very careful not to touch the profilometer tip, as this is very delicate) 
  • Turn on the PLX-Benchtop, open Kinesis and 'Home' the stages
  • Close Kinesis, open CORSICA and try to run a test

If the issue persists or an obstruction was felt during the moving of the arm, please contact support